Regular field adjustments ensure your digital level maintains its accuracy and reliability. This guide covers essential adjustments you can perform in the field to maintain optimal performance.
Always perform adjustments in stable conditions and verify results with redundant measurements.
1. Set up the instrument on a stable surface
2. Center the bubble using leveling screws
3. Rotate instrument 180°
4. If bubble moves off center, adjust halfway using adjustment screws
5. Repeat until bubble stays centered through full rotation
1. Set up two points exactly 60m apart
2. Place level exactly in middle (30m from each point)
3. Take readings on both points (A1, B1)
4. Move instrument to 2m behind point A
5. Take new readings (A2, B2)
6. Calculate adjustment value: (B2 - A2) - (B1 - A1)
1. Access calibration menu in instrument settings
2. Follow manufacturer's calibration sequence
3. Verify calibration with known reference points
4. Document calibration results and date